Generator constant potential (high voltage DC, 100 – 200 kHz)
Tube voltage 60 or 70 kV
Tube current 7 mA
Focal spot 0.7 mm to IEC 60336
Total filtration 2.5 mm Al (70 kV)
Line voltage 115 / 230 VAC +/- 10 %, 50/60 Hz
Exposure time range 0.02 – 3.2 seconds
Expandability
Remote exposure button
Two remote controls
Door switch readiness
External exposure lamp readiness
Unit mount
Weight Total 66 lbs (30 kg)
Beam limiting cone dimensions Round diameter 2.36” (60 mm), rectangular 1.38” x 1.77” (35 x 45 mm)
SSD (Source-Skin Distance) Standard cone 9” (229 mm), long cone 12” (305 mm)
Installation Vertical wall mount for single stud, or horizontal wall mount for 16” (406 mm) center stud installation unit mount

ISO9001 Quality Certified: Inline Systems is committed to providing professional and superior service.

ISO9001 Quality Certified
Inline Systems is committed to providing professional and superior service.